Summary
A survey of how audiologists evaluate hearing aid telecoils, with a proposed protocol for setting the telecoil programme, from a themed issue on telephones and telecoils.
Key findings
- Reports survey data showing no clear pattern in how audiologists evaluate hearing aids with telecoils; the great majority relied on informal patient judgments of sound quality rather than probe-microphone or speech-recognition measures.
- Reviews formal telecoil-fitting protocols and finds that none has gained wide acceptance, perhaps because of the time and equipment they need.
- Argues that protocols using an induction loop input do not represent telephone use, and proposes a seven-step protocol that uses a standard inductive signal source to play a known speech or noise signal while the telecoil memory is programmed.
Limits
Read from the publisher's abstract only; the full text, including the survey's size, method and date, could not be read. The article is about telephone use and dates from 2003. Its authors worked for a hearing aid maker and Pehratek Products.
Read the original
www.thieme-connect.com/products/ejournals/abstract/10.1055/s-2003-37909
Cited in
Cite this source
Yanz JL, Pehringer J. Quantifying Telecoil Performance in the Ear: Common Practices and a New Protocol. Seminars in Hearing 24(1):71-80 (Thieme). 13 March 2003. https://www.thieme-connect.com/products/ejournals/abstract/10.1055/s-2003-37909 (accessed 9 October 2026).
Every finding here cites its source, and who funded each source is labelled. This site does not recommend products. Found an error or a missing study? Suggest a correction.