Summary
A study comparing a directly connected FM receiver with a neckloop into the telecoil of two Cochlear implant processors, for speech understanding in quiet and in noise.
Key findings
- Adults with at least a year's implant use (15 in the first experiment, 16 in the second) used Cochlear's Nucleus Freedom and Nucleus 5 processors with a directly connected (direct audio input) FM receiver and with an induction neckloop receiver.
- Speech recognition in quiet did not differ between the two FM systems or between the processor conditions.
- With each processor, speech recognition in noise was better with the directly connected receiver than with the neckloop.
- The Nucleus 5's multistage automatic sensitivity control gave better results than the single-stage approach; the authors conclude that speech recognition in noise depends substantially on the processor, the FM system and how automatic sensitivity control is applied.
Limits
Read from the publisher's abstract only; the full text could not be read. Personal FM receivers, not venue systems; one implant maker's processors of 2013.
Read the original
Cited in
- Are telecoils switched on and set up well enough for loops to work?Not reliable yet
- What reaches a cochlear implant user in a venue, and how well?Partly answered
- Do neckloops give telecoil users good sound from FM, infrared, Wi-Fi and Auracast receivers?Not reliable yet
Cite this source
Wolfe J, Schafer E, Parkinson A, John A, Hudson M, Wheeler J, Mucci A. Effects of Input Processing and Type of Personal Frequency Modulation System on Speech-Recognition Performance of Adults With Cochlear Implants. Ear and Hearing 34(1):52-62 (Lippincott Williams & Wilkins). January 2013. https://www.ovid.com/jnls/ear-hearing/abstract/10.1097/aud.0b013e3182611982~effects-of-input-processing-and-type-of-personal-frequency (accessed 9 October 2026).
Every finding here cites its source, and who funded each source is labelled. This site does not recommend products. Found an error or a missing study? Suggest a correction.