Neckloop measurements
Neckloop and receiver-pair results, with field or level observations, frequency response, test conditions and reading limits.
Data licence: CC0 1.0 (public domain dedication).
Reading the results
Rows retain named models or receiver-neckloop pairs, alongside explicitly labelled summaries whose individual models are unavailable. Input voltage, magnetic field strength and relative frequency-response plots measure different things. Read the reference level, position and driving source with each result. Abstract-only findings retain their reading limits; personal-device paths concern shared coupling rather than venue performance. (Pro Audito Schweiz 2026; Hallikainen 2019; Hawkins and Schum 1985; Schafer et al. 2007)
Column definitions · Compilation and interpretation · Change log
Scroll horizontally to read all columns. Empty cells mean no value was recorded; read each row’s notes and limits with its results.
| Result ID | Model | Maker | Receiver or source | Row scope | Field strength or level | Input level | Frequency response note | Distortion or noise note | Test method | Conditions | Sample size | Test date | Source date | Source ID | Source locator | Funding detail | Scope | Limitations | Funding |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| pro-audito-2026-neckloops-iloop-plus | iLoop+ | Geemarc | Bench drive not identified; no named receiver or phone output | model or pair | not reported | not reported | Most nearly flat high-frequency response of the six models; reduced low-frequency response | not reported | Pro Audito reports dummy-head measurements according to IEC 62489-1; relative frequency-response plot referenced to 1 kHz | Receiver output and dummy-head geometry not detailed; no separate phone-output result set. Active loop with integrated amplifier and battery | Six models in comparison; units per model not reported | not reported | 2026-09-08 | pro-audito-2026-neckloops | English reprint pp. 2, 6 | Independent (Pro Audito Schweiz, Swiss hearing-loss association) | venue component | English translation of German report. Project funding not stated. No raw frequency samples or quantified distortion/noise; graph not digitized. Non-linear headphone outputs may change the response. | Independent |
| pro-audito-2026-neckloops-la-438 | LA-438 | Listen Technologies | Bench drive not identified; no named receiver or phone output | model or pair | 400 mA/m (0 dB re 400 mA/m) | -12.7 dBu for the reported reference field | High-frequency roll-off, less pronounced than in the other passive models | Authors warn of distortion at maximum volume on weak headphone outputs; no distortion percentage reported | Pro Audito reports dummy-head measurements according to IEC 62489-1; relative frequency-response plot referenced to 1 kHz | Receiver output and dummy-head geometry not detailed; no separate phone-output result set. Passive loop; reported impedance 3-6 ohms at 1 kHz | Six models in comparison; units per model not reported | not reported | 2026-09-08 | pro-audito-2026-neckloops | English reprint pp. 3, 6 | Independent (Pro Audito Schweiz, Swiss hearing-loss association) | venue component | English translation of German report. Project funding not stated. No raw frequency samples or quantified distortion/noise; graph not digitized. Non-linear headphone outputs may change the response. | Independent |
| pro-audito-2026-neckloops-il-200 | IL 200 | Beyerdynamic / Televic | Bench drive not identified; no named receiver or phone output | model or pair | 400 mA/m (0 dB re 400 mA/m) | -6.4 dBu for the reported reference field | Marked high-frequency loss | not reported | Pro Audito reports dummy-head measurements according to IEC 62489-1; relative frequency-response plot referenced to 1 kHz | Receiver output and dummy-head geometry not detailed; no separate phone-output result set | Six models in comparison; units per model not reported | not reported | 2026-09-08 | pro-audito-2026-neckloops | English reprint pp. 4, 6 | Independent (Pro Audito Schweiz, Swiss hearing-loss association) | venue component | English translation of German report. Project funding not stated. No raw frequency samples or quantified distortion/noise; graph not digitized. Non-linear headphone outputs may change the response. | Independent |
| pro-audito-2026-neckloops-ezt-3012 | EZT-3012 | Sennheiser | Bench drive not identified; no named receiver or phone output | model or pair | 400 mA/m (0 dB re 400 mA/m) | +0.3 dBu for the reported reference field | Marked high-frequency loss | not reported | Pro Audito reports dummy-head measurements according to IEC 62489-1; relative frequency-response plot referenced to 1 kHz | Receiver output and dummy-head geometry not detailed; no separate phone-output result set | Six models in comparison; units per model not reported | not reported | 2026-09-08 | pro-audito-2026-neckloops | English reprint pp. 4, 6 | Independent (Pro Audito Schweiz, Swiss hearing-loss association) | venue component | English translation of German report. Project funding not stated. No raw frequency samples or quantified distortion/noise; graph not digitized. Non-linear headphone outputs may change the response. | Independent |
| pro-audito-2026-neckloops-nl-90 | NL-90 | Monacor | Bench drive not identified; no named receiver or phone output | model or pair | not reported | not reported | Very strong high-frequency roll-off | not reported | Pro Audito reports dummy-head measurements according to IEC 62489-1; relative frequency-response plot referenced to 1 kHz | Receiver output and dummy-head geometry not detailed; no separate phone-output result set | Six models in comparison; units per model not reported | not reported | 2026-09-08 | pro-audito-2026-neckloops | English reprint pp. 5, 6 | Independent (Pro Audito Schweiz, Swiss hearing-loss association) | venue component | English translation of German report. Project funding not stated. No raw frequency samples or quantified distortion/noise; graph not digitized. Non-linear headphone outputs may change the response. | Independent |
| pro-audito-2026-neckloops-nkl-001s | NKL-001S | Williams AV | Bench drive not identified; no named receiver or phone output | model or pair | not reported | not reported | Very uneven response; rising impedance with frequency makes the response depend strongly on the source. Authors found no acceptable result with the headphone outputs tried | not reported | Pro Audito reports dummy-head measurements according to IEC 62489-1; relative frequency-response plot referenced to 1 kHz | Receiver output and dummy-head geometry not detailed; no separate phone-output result set | Six models in comparison; units per model not reported | not reported | 2026-09-08 | pro-audito-2026-neckloops | English reprint pp. 5, 6 | Independent (Pro Audito Schweiz, Swiss hearing-loss association) | venue component | English translation of German report. Project funding not stated. No raw frequency samples or quantified distortion/noise; graph not digitized. Non-linear headphone outputs may change the response. | Independent |
| hallikainen-2019-neckloop-notes-ntc-102 | NTC-102 | Comtek | IRH-230 infrared headset receiver | model or pair | >19 dB re 100 mA/m at loop centre (author-inferred full-volume lower bound) | Receiver at full volume; no-carrier noise, with no infrared signal present | not reported | No distortion or signal-to-noise ratio reported. Receiver no-carrier noise was the test stimulus | Manufacturer engineer used FoSmeter+ vertically at centre of neckloop lying flat on table. Reduced volume until unmarked loop read 0 dB, then compared loops at that setting | +3 dB relative to unmarked loop at reduced drive; approximate impedance 36 ohms, measured at 400 Hz through 12-ohm series resistor | Four neckloops identified; repeat count not reported | 2012 | 2019-01-26 | hallikainen-2019-neckloop-notes | Hallikainen reply at 11:32 AM; Hearing Aid Compatibility and field-strength table | Manufacturer (Harold Hallikainen's cinema-receiver test notes; the author describes the tested headphones as their own; neckloops and field-strength meter provided by Connections Unlimited; project funding not stated) | venue receiver and neckloop | Informal Manufacturer test notes. Centre readings exceeded meter range; >19 dB is inferred from +3 dB relative to the unmarked loop, whose full-volume field exceeded +16 dB. Table geometry does not reproduce a telecoil at the ear. Exact date of centre measurements not reported; notes place testing in 2012. | Manufacturer |
| hallikainen-2019-neckloop-notes-noizfree | NoizFree (no model number stated) | not reported | IRH-230 infrared headset receiver | model or pair | >19 dB re 100 mA/m at loop centre (author-inferred full-volume lower bound) | Receiver at full volume; no-carrier noise, with no infrared signal present | not reported | No distortion or signal-to-noise ratio reported. Receiver no-carrier noise was the test stimulus | Manufacturer engineer used FoSmeter+ vertically at centre of neckloop lying flat on table. Reduced volume until unmarked loop read 0 dB, then compared loops at that setting | +3 dB relative to unmarked loop at reduced drive; approximate impedance 27 ohms, measured at 400 Hz through 12-ohm series resistor | Four neckloops identified; repeat count not reported | 2012 | 2019-01-26 | hallikainen-2019-neckloop-notes | Hallikainen reply at 11:32 AM; Hearing Aid Compatibility and field-strength table | Manufacturer (Harold Hallikainen's cinema-receiver test notes; the author describes the tested headphones as their own; neckloops and field-strength meter provided by Connections Unlimited; project funding not stated) | venue receiver and neckloop | Informal Manufacturer test notes. Centre readings exceeded meter range; >19 dB is inferred from +3 dB relative to the unmarked loop, whose full-volume field exceeded +16 dB. Table geometry does not reproduce a telecoil at the ear. Exact date of centre measurements not reported; notes place testing in 2012. | Manufacturer |
| hallikainen-2019-neckloop-notes-ezt1011 | EZT1011 | Sennheiser | IRH-230 infrared headset receiver | model or pair | >19 dB re 100 mA/m at loop centre (author-inferred full-volume lower bound) | Receiver at full volume; no-carrier noise, with no infrared signal present | not reported | No distortion or signal-to-noise ratio reported. Receiver no-carrier noise was the test stimulus | Manufacturer engineer used FoSmeter+ vertically at centre of neckloop lying flat on table. Reduced volume until unmarked loop read 0 dB, then compared loops at that setting | +3 dB relative to unmarked loop at reduced drive; approximate impedance 33 ohms, measured at 400 Hz through 12-ohm series resistor | Four neckloops identified; repeat count not reported | 2012 | 2019-01-26 | hallikainen-2019-neckloop-notes | Hallikainen reply at 11:32 AM; Hearing Aid Compatibility and field-strength table | Manufacturer (Harold Hallikainen's cinema-receiver test notes; the author describes the tested headphones as their own; neckloops and field-strength meter provided by Connections Unlimited; project funding not stated) | venue receiver and neckloop | Informal Manufacturer test notes. Centre readings exceeded meter range; >19 dB is inferred from +3 dB relative to the unmarked loop, whose full-volume field exceeded +16 dB. Table geometry does not reproduce a telecoil at the ear. Exact date of centre measurements not reported; notes place testing in 2012. | Manufacturer |
| hallikainen-2019-neckloop-notes-unmarked | Unmarked neckloop | not reported | IRH-230 infrared headset receiver | model or pair | Centre: >16 dB re 100 mA/m with full-volume no-carrier noise; 6 inches above loop: +3 dB re 100 mA/m with full-volume no-carrier noise; 6 inches outside loop: +12 dB re 100 mA/m with programme audio | Centre and above-loop readings: receiver full-volume no-carrier noise; outside-loop reading: programme audio through IRC-28C infrared panel | not reported | No distortion or signal-to-noise ratio reported. No-carrier noise was the stimulus for the centre and above-loop readings; programme audio was used outside the loop | Manufacturer engineer used FoSmeter+ vertically at centre of neckloop lying flat on table. Reduced volume until unmarked loop read 0 dB, then compared loops at that setting | Loop flat on table for centre and above-loop readings. Approximate impedance 34 ohms at 400 Hz through 12-ohm series resistor. Above-loop and outside-loop notes dated 2012-08-13 | Four neckloops identified; repeat count not reported | 2012 | 2019-01-26 | hallikainen-2019-neckloop-notes | Hallikainen reply at 11:32 AM; field-strength table and note ending hh 8/13/12 | Manufacturer (Harold Hallikainen's cinema-receiver test notes; the author describes the tested headphones as their own; neckloops and field-strength meter provided by Connections Unlimited; project funding not stated) | venue receiver and neckloop | Informal Manufacturer test notes. Centre value is a lower bound beyond the meter range. Above and outside are distinct stated positions, combined here only to retain one row per receiver-neckloop pair. No ear-position measurement. The two unmarked entries in the source table describe the reference loop, not two identified models. | Manufacturer |
| hawkins-schum-1985-neckloop-summary | not reported | not reported | Two unidentified personal FM systems coupled to four hearing aids | study summary; models unavailable | not reported | not reported | Some coupling comparisons had large response differences; no coupling method consistently matched the hearing-aid-alone response best. Neckloop-specific curves unavailable | Distortion and noise were similar for hearing aids alone and FM-hearing-aid combinations except with the silhouette inductor | Electroacoustic measurement of frequency response, harmonic distortion, noise, input-output functions and receiver volume-wheel taper; compared hearing aids alone with direct input, neckloop and silhouette coupling | Some FM-hearing-aid combinations had strongly non-linear volume-wheel curves; abstract does not identify those pairs | Four hearing aids and two FM systems; number of neckloops not reported | not reported | 1985-05 | hawkins-schum-1985 | Archived public abstract (PMID 3990259), methods and results | Independent (journal study; the archived abstract and metadata give no author affiliation or funder, and the full text was not read) | shared-path evidence | Abstract only. Personal classroom FM equipment, not venue receivers. Models, magnitudes and individual pair results unavailable. This is one pooled finding, not one identified model or every possible device combination. Funder not stated in abstract. | Independent |
| schafer-2007-neckloop-summary | not reported | not reported | Unidentified body-worn personal FM receiver(s) coupled to hearing aids via neckloop | study summary; models unavailable | not reported | not reported | Significant low-frequency reduction; greatest variability of the three receiver types | not reported | ASHA 2002 FM-advantage method; compared hearing-aid-alone and FM-system outputs at typical input levels using a complex signal and at high input levels using a pure tone | Compared body-worn neckloop, nonprogrammable ear-level and programmable ear-level receivers | Three receiver types; tested unit and hearing-aid counts unavailable in abstract | not reported | 2007-10 | schafer-2007 | Archived public abstract (PMID 17890512), Method and Results | Independent (first author at the Speech and Hearing Sciences Department, University of North Texas; the PubMed record names no funder, and the full text, which may state one, was not read) | shared-path evidence | Abstract only. Personal FM fitting evidence, not venue receiver measurements. Models, numerical input/output levels and size of frequency loss unavailable. Funder not stated in abstract. | Independent |
| salzbrenner-1991-pocketalker-neckloop | Induction neckloop (model not stated) | not reported | Williams Sound PockeTalker personal amplifier; neckloop coupled to wearer telecoil | model or pair | not reported | not reported | Across tested devices, non-telecoil coupling gave broader and flatter responses; average telecoil versus non-telecoil differences were minimal. Both PockeTalker versions showed gain variation across samples at all frequencies | not reported | Madsen IGO 1000 in-situ probe-tube measurements of product uniformity, volume-control linearity, ear-coupling resonance effects and frequency response | not reported | One wearer; three samples of each of six device configurations, including PockeTalker with neckloop and permanently wired PockeTalker | not reported | 1991-04 | salzbrenner-1991 | Archived public abstract (PMID 1829011) | Independent (Cleveland Hearing and Speech Center, Ohio, according to the PubMed record; no funder is named there) | shared-path evidence | Abstract only. Personal amplifier path, not a venue installation. Neckloop maker and model not stated; Williams Sound identifies the amplifier maker only. No numerical response differences or individual sample readings. Broader/flatter finding is a pooled comparison, not an isolated neckloop curve. Funder not stated in abstract. | Independent |
| rootlieb-2024-roger-neckloop-nucleus7 | Roger Neckloop | Phonak | Apple MacBook Air source; Cochlear Nucleus 7 processor in telecoil mode | model or pair | Absolute levels unavailable: spectra aligned at 1 kHz; 65 dBA was a setup calibration target, not a measured neckloop field | not reported | Authors report generally less than 5 dB spectral differences above 500 Hz across wireless routes versus loudspeaker, with larger differences in some lower bands. Figure 1 includes the telecoil route; this is not a strict bound for every frequency | No quantified distortion or noise result for neckloop. Discussion notes magnetic-interference sensitivity | Speech-shaped masking noise through processor listening-check output, Samson Servo 120a amplifier and Sennheiser HDA 200 headphones on Bruel & Kjaer 4153 artificial ear with 2250 sound level meter; third-octave spectra | Neckloop suspended around processor; processor microphone and optional preprocessing disabled. Laptop substituted because smartphone could not connect. Placement affected volume | One processor model in telecoil condition; repeat unit count not stated. Separate listening experiment: 20 normal-hearing adults | not reported | 2024-08-05 | rootlieb-2024 | Printed pp. 315-316, Experiment 1 Methods/Results, Table 1 and Figure 1; p. 318 Discussion | Independent (Amsterdam UMC; no funder is named and the authors report no conflict of interest) | shared-path evidence | Remote self-testing laboratory setup with personal devices, not venue equipment or implant-user listening. Readable full article with repository cover. Monitor-output spectra do not measure magnetic field or absolute implant output; no graph values digitized. Article names no funder. Source date is first online publication; issue is April 2025. | Independent |
| pro-audito-2025-operator-guidance-unnamed-pairs | not reported | not reported | Unidentified FM, infrared and Auracast transmitter-receiver-neckloop pairs | study summary; models unavailable | Required field at ear reached only with full transmitter drive and receiver volume usually at maximum; numerical measured field not reported | 1 kHz sine wave; transmitter fully driven; receiver volume usually at maximum | not reported | not reported | Operator guidance reports magnetic-field measurements at the ear using a 1 kHz sine wave; meter and geometry not stated | not reported | not reported | not reported | 2025-02-17 | pro-audito-2025-operator-guidance | Section on magnetic field strength at the neckloop | Nonprofit advocacy (Pro Audito Schweiz, Swiss hearing-loss association) | venue receiver and neckloop | Unidentified pooled systems; cannot be assigned to specific models or counted as individual tested pairs. Guidance page has an unstated revision history, so its displayed date need not date the measurements. No raw readings or detailed protocol. | Nonprofit advocacy |
No matching rows.
Codebook
result_idUnique source id plus a model or result suffix.
modelNeckloop model as identified by the source. An unidentified loop stays unidentified; the model of its amplifier is in
receiver_or_source.makerNeckloop maker only. A source that identifies the amplifier maker does not necessarily identify the neckloop maker.
receiver_or_sourceDriving receiver or audio source, with the receiving hearing device where relevant.
row_scopemodel or pair, orstudy summary; models unavailablefor pooled results that cannot be assigned to models.field_strength_or_levelReported magnetic field, qualified field lower bound, or explanation of why an absolute level is unavailable. Units, reference level and position remain with each value.
input_levelInput voltage or receiver drive condition. dBu describes electrical input, not magnetic field strength.
frequency_response_noteMeasured response in the source's own comparison context, paraphrased. Pooled findings and normalized spectra are identified.
distortion_or_noise_noteMeasured comparison, qualitative source warning, or absence of a reported result. A warning is not a measured distortion percentage; test noise is not a noise-performance result.
test_methodInstrument, signal, coupling route, reference and measurement procedure where stated.
conditionsPosition, drive, impedance, processing and other setup details that affect interpretation.
sample_sizeDevices, configurations or wearers where known; a separate listener study is identified separately. Unreported replication is not assumed to be one unit.
test_dateMeasurement date at supported precision, or
not reported. Publication dates are never substituted. The cinema notes place testing in 2012; the unmarked-loop row also identifies the two explicitly dated 13 August readings inconditions.source_datePublication or displayed source date, using
YYYY-MM-DDorYYYY-MM. The operator-guidance page has an unstated revision history.source_idLibrary source identifiers supporting the row, separated by semicolons. Feature-specific and supporting-source columns retain their more precise roles.
source_locatorOriginal page, section, table or model entry supporting the row; source names identify locations when several originals are used.
funding_detailFunding and producer context retained from the earlier compilation; the funding column carries the current source-record labels.
scopevenue component,venue receiver and neckloop, orshared-path evidence. Personal FM fitting, personal amplification and remote implant testing inform the shared coupling path, without establishing venue performance.limitationsRow-specific interpretation limits and source-record reading limits. Missing information is not reported, never an inferred zero.
fundingSource-record funding labels. With multiple primary sources, each label is followed by its source id in brackets. Funding detail remains in funding_detail where supplied by the earlier compilation.
Compilation and interpretation
neckloops.csv was compiled and independently audited on 2026-10-13 from seven existing library records and their private archived originals. It contains 15 rows: six neckloop models from Pro Audito's dummy-head test, four receiver-neckloop pairs from Hallikainen's Manufacturer test notes, two personal-device paths, and three explicitly labelled study summaries whose individual models cannot be recovered from the readable material.
One row retains the results for a neckloop model or receiver-neckloop pair within one source. Multiple measurement positions for the same pair stay together with their individual conditions. The three study summary; models unavailable rows are exceptions to model-level granularity: they preserve the 1985 and 2007 abstract findings and Pro Audito's unnamed system tests without inventing device identities or expanding pooled results into individual pairs. Do not count those rows as three tested products.
The CSV is UTF-8, comma-delimited, with one header, double-quoted fields and LF line endings. Import all fields as text to preserve identifiers, partial dates, inequality signs and mixed measurement units. not reported means unavailable in the material read, never zero. The file contains no formulas, conversions, rankings or digitized graph points. Rows are grouped by source rather than performance.
The measurement types cannot be pooled into a product average. Pro Audito's three explicit sensitivity results give the electrical inputs for 400 mA/m; its frequency-response graph is relative to 1 kHz. The cinema notes use dB relative to 100 mA/m, with loops on a table: the three >19 dB values are the author's inferred full-volume lower bounds from a reduced-volume comparison after the reference loop exceeded the meter's range. The unmarked loop's centre, above-loop and outside-loop readings are different conditions. Rootlieb's processor-output spectra were aligned at 1 kHz, so their plotted absolute levels cannot supply a neckloop field or absolute device output. (Pro Audito Schweiz 2026; Hallikainen 2019; Rootlieb et al. 2024)
The 1985, 1991 and 2007 papers are represented only by archived public abstracts and metadata. Missing full-text models and numerical differences remain unavailable. The table preserves both adverse response findings and the reports of similar distortion/noise or small average coupling differences. Specifications without a reported test, standards targets, speech-recognition outcomes, general user impressions, latency and battery duration are outside this measurement table. The q22 research report lists screening decisions, archive provenance and existing-record concerns. (Hawkins and Schum 1985; Salzbrenner et al. 1991; Schafer et al. 2007)
For updates, read the source record and archived original first. Keep results from different sources or test conditions distinguishable. Replace a pooled summary with individual pair rows only when the underlying results become readable, retaining its provenance and documenting the change.
Change log
- 2026-10-13: Migrated 15 rows with codebook and source provenance into the data section.
Cite this table
Equal Access Audio. “Neckloop measurements” Assistive Listening Knowledge Base. Version 1.0. Compiled 13 October 2026; updated 13 October 2026. https://assistivelisteningknowledgebase.org/data/neckloops/ (accessed the date you read it).